Universal Loader - Operation Instructions
|
Z:\Quantum \Production Documentation\SP-016b.doc
1/25/02
Model-555/9500 System Software Upgrade Instructions
Please perform the following steps as instructed.
If you encounter any problems you can contact us at:
Phone: 800-510-6530
Fax: 406-582-0237
Email:
productsupport@quantumcomposers.com
Mail: Quantum Composers
P.O. Box 4248
Bozeman, MT 59772
USA
System Requirement:
·
Computer :
486 Processor or higher
·
Operating System : XP, Windows 95/98, or Windows NT
Hardware Setup:
·
Your PC should have a 9-pin or a 25-pin connector on the back that is labeled
COM1, COM2, COM3, or COM4. (Sometimes the ports are labeled COMA,
COMB...)
·
You will need to attach a cable from the COM port on your PC, to the connector
labeled RS232 on the back of the instrument. The cable should be a straight-thru
connection, that is, pin 1 of one end is connected to pin 1 of the other and so on.
Do NOT use a null-modem cable.
·
A 9-pin to 9-pin cable is normally provided with your instrument. A 9-pin to 25-
pin adapter may be required if your PC has a 25-pin connector. These adapters
can be obtained at most computer stores.
Z:\Quantum \Production Documentation\SP-016b.doc
1/25/02
Universal Loader Operation Instructions:
Note: It is very important that you read all instructions displayed during the application
carefully. Failure to do so may cause the instrument to be set up incorrectly, resulting in
erroneous operation of the instrument.
1)
Connect the RS232 cable to your instrument.
2)
Turn instrument OFF.
3)
Launch uploader.exe application and read initial screen carefully.
4)
Select serial port for communications with the instrument.
5)
Find file containing firmware update data. Select the file name that matches
the model number printed on the back of your instrument. The version
number of the file name will vary depending on the revision number of the
firmware.
6)
Set upload reset level. This function is for engineering use only. Never check
the "Reset All Variables" box unless you are instructed to do so.
7)
Power up the instrument when instructed to do so by the application. If
"Pulse Generator not responding" message is display a communications
timeout has occurred. Verify serial cable connection and communication port
selection. Try turning the instrument OFF then ON.
8)
When communications with the instrument is verified by the application. The
application will transition to the "Transfer Update" screen. The Status
Window will indicate "Ready to receive update". Press the UPDATE button
at this time.
9)
The upload process will proceed by displaying the upload progress in both the
STATUS Window and on the instruments display.
10)
The upload process is complete when the "Update Completed!" message box
is displayed. Click OK to clear this message box.
11)
Wait for the instrument to emit two beeps and returned to the "Setup" display
("Setup" will be displayed in the upper right hand corner of the display)
12)
Power down the instrument.
13)
Following a successful upload the application will return to the "Locate File"
screen. You may upload the firmware in another instrument at this time or
exit the application.
|